| | Product Group: | Calibrators |
| Product Category: | Hand-held |
| Product Name: | Temperature calibrator |
| Product Code: | Calog-TEMP |
| Brochure: | 376 Kb |
| Short Description: | Thermocouple and RTD calibrator with measurement and source of millivolts, ohms and mA's plus logging
|
|
|
|
Calog-TEMP is a portable calibrator which is made for testing, measuring and calibrating all Thermocouple and RTD based systems. Its small size, rugged design, rubber protective boot, high precision and long-lasting re-chargeable lithium ion battery pack makes it well suited to industrial field calibration.
Using the simultaneous, isolated, combination of Thermocouple or RTD sourcing and mA measuring it is possible to calibrate any Tc/mA or RTD/mA temperature transmitter. The mA measurement can be passive with a remote supply or active 4-20mA 2-wire with the Calog-TEMP generating the 24VDC loop supply.
The accurate mA measure and source on the calibrator allows it to be used for all mA loop testing and instrument calibration. For temperature monitoring over time Calog calibrators may be set in trend mode with a selectable timebase to graph a temperature profile. This is ideal for testing ovens or for optimising PID temperature controllers. Data may be logged to the removable SD card in CSV format, with user-selectable data fields and timebase. Logged data may also be retrieved via the USB port.
Features:
- Small, rugged, handheld with a protective rubber cover
- Graphic display of measured value, percent and battery status
- Measure, source and simulate 0-24 mA with 0.01% accuracy and 1 microamp resolution
- Continuity test function
- Trend logging with programmable timebase
- Datalogging to SD card with programmable fields and times
- USB port for access to logged data
- Displays values in mA and % or % and mA
- Programmable auto-off, restart at last setting and selectable display resolution
- Measure and source types K, J, T, B, R, S, E, N, U and L thermocouples
- Configurable internal, external or manual cold-junction compensation
- Accuracy better than 0.01% of FS excluding CJ errors and Tc errors with 0.1°C resolution
- Measure and source Pt50, Pt100, Pt500, Pt1000, Ni100 and Ni120
- Simultaneously source thermocouples or RTD's and measure mA
- Measure and source Ohms and RTD in 2, 3 or 4-wire configurations
- Measure and source -10 to 100 mV with 0.01% accuracy and 1 microvolt resolution
- Measure and source 0 to 24mA with 0.02% accuracy and 1 microamp resolution
- Displays in °C, °F, Kelvin or Rankine
- Lithium Ion battery pack with 'fuel gauge' chip
- Built-in 24V transmitter power supply
- Plug-on bare-wire Tc CJC module supplied as standard
- Carry case, battery charger and test leads supplied as standard
- 2 year material and workmanship guarantee (excludes batteries)
|
|
Technical Specifications: Temperature Calibrator
|
| Analogue Input Ranges |
Impedance |
Accuracy |
Resolution |
| 0 to 24 mA |
Input Impedance ± 1 Ω, 250 mA Fused |
0.01 % |
1 µA |
| -10 to 100 mV |
Input Impedance > 1 MΩ |
0.01 % |
1 µA |
| Tc Types K, J, T, B, R, S, E, N, U and L |
Input Impedance > 1 MΩ |
0.01 %* |
0.1 °C |
| 0 - 400.00 Ω |
|
0.05 % |
0.01 Ω |
| 0 - 2000.0 Ω |
|
0.05 % |
0.1 Ω |
| Pt50, Pt100, Pt500, Pt1000, Ni100, Ni120 |
|
0.05 % |
0.01 °C |
Measurement of Ohms and RTD's in 2, 3 or 4-wire configurations
|
Continuity with 100 Ω trigger and 1mA excitation, with visual and audible confirmation
|
| Analogue Output Ranges |
Load |
Accuracy |
Resolution |
| Thermocouple Types K, J, T, B, R, S, E, N, U and L |
0.01 %* |
0.1 °C |
| 0 to 24 mA |
Output load max 500 Ω |
0.01 % |
1 µA |
| -10 to 100 mV |
Output load min 600 Ω |
0.01 % |
1 µV |
RTD types Pt50, Pt100, Pt500, Pt1000, Ni100, Ni120 |
0.05 % |
0.01 °C |
| 1 to 400.00 Ω |
10 to 400 Ω |
0.05 % |
0.01 Ω |
| 1 to 2200.0 Ω |
10 to 2k2 Ω |
0.05 % |
0.1 Ω |
| |
|
* excluding cold junction |
|
|
|
|
|