Pressure Calibrator - Calog-Pressure II

Pressure and mA calibrator
CALOG-Pressure II is a handheld calibrator with plug-on pressure sensors for any range up to 1000 bar. This instrument also has the functions to install and maintain all powered and non-powered milliamp loops. It is supplied with 1 plug-on pressure sensor with optional additional sensors available for the other pressure ranges specified below.

As a small, rugged unit with rubber protective boot, high precision and Li-ion long-lasting rechargeable battery pack, CALOG-Pressure II is ideal for industrial field calibration. Data may be logged to the removable SD card in CSV format, with user-selectable data fields and timebase. Logged data may also be retrieved via the USB port. CALOG calibrators are made by Instrotech and comply with CE and C-Tick EMC and EMI and Low Voltage directives.
  • Small, rugged, handheld with a protective rubber cover
  • Graphic display of measured value, percent and battery status
  • Measure, source and simulate 0-24 mA with 0.01% accuracy and 1 microamp resolution
  • Measure 32V with 0.005% accuracy and 1mV resolution
  • Continuity test function
  • Trend logging with programmable timebase
  • Displays values in mA and % or % and mA
  • Measure complete vacuum to 1000 bar pressure
  • Measures in bar, mbar, kPa, MPa, psi, m of H²O, in of H²O, ft of H²O, mm of Hg (torr), cm of Hg, in of Hg, kg per cm² and atmospheres
  • Contains a handy pressure conversion calculator
  • Programmable auto-off, restart at last setting and selectable display resolution
  • Data logging to removable SD card with time and date
  • USB port for access to logged data
  • Lithium-ion rechargeable battery pack with 'fuel gauge' chip
  • One pressure sensor included, more are optional
  • Carry case, battery charger, SD card, test leads and factory calibration certificate
  • 2 year guarantee, excluding batteries
Pressure Measurement Ranges Accuracy Resolution
-1 to 1 bar, 0 to 1 / 3 / 10 / 30 / 100 / 300 / 700 bar Special ranges on request 0.05 % of FS Sensor dependent
Analogue Inputs Impedance Accuracy Resolution
0 to 24 mA Input Impedance ± 17 Ω, 250mA fused 0.01 % of FS 1 µA
0 to 32 V Input Impedance ±110 kΩ 0.005 % of FS 1 mV
Continuity with a <100 Ω trigger and 1 mA excitation with visual and audible confirmation
Analogue Output Load Accuracy Resolution
0 to 24 mA Output load max 500 Ω 0.01 % of FS 1 µA

Calog-Press II pressure calibrator

  • Reload
Pressure Calibrator
Calog Pressure Calibrator Kit
Pressure Calibrator display
Pressure Calibrator display
Pressure Calibrator Trend Graph
Calog Pressure Calibrator Trend Graph
Calog pressure calibrator functions
Calog pressure calibrator functions

Copyright © 1999-2017 Instrotech Australia Pty Ltd. All rights reserved.
E&OE. Pricing and Availability may be subject to change without notice.